Chapter 5 Maintenance and Storage
88
Chap. 5
Maintenance and Storage
1.3
Cleaning Immersion Oil
(1) Wipe with petroleum benzine.
(2) Finish off the cleaning with absolute alcohol (ethyl or methyl alcohol) after cleaning with petroleum benzene.
If petroleum benzene is not available
If petroleum benzene is unavailable, use methyl alcohol alone. However, typically wipe three or four times
because the detergency is weak.
1.4
Decontaminating the Product
For routine disinfection of this product, Nikon recommends using 70% medical alcohol.
Use of organic solvents on plastic parts may result in discoloration.
Cautions on disposal
If contact occurs between a sample and this product, determine whether the sample is hazardous. If the sample
is hazardous, follow the standard procedures for your facility.
2
Storage
Store this product in a dry location where mold is unlikely to form.
Storage conditions are as follows: temperature (-20°C to +60°C), humidity (90% RH max., no condensation)
Store the objectives and eyepieces in a desiccator or similar container with a drying agent.
Place a cover over this product to protect it from dust.
Switch off the microscope (press the switch to the “O” position) and wait for the lamphouse to cool before covering this
product with a cover.
3
Regular Inspections (Charged)
To maintain the performance of this product, Nikon recommends periodic inspections (chargeable service). Contact your