Chapter 4 Troubleshooting
86
Chap. 4
Troubleshooting
■ Capture button
Problem Cause Measure
The capture button does
not function.
A camera trigger cable is not properly
connected.
Attach it correctly.
(→Chapter 3, “7 Attaching a Camera”.)
A destination of a capture trigger signal has
not been set correctly.
Change the setting with DS-L4 or Ni Setup
Tool.
2.2
Epi-fluorescence Microscopy
Problem Cause Measure
The mercury lamp does
not work.
There is no power supplied.
Plug in the power cord.
(→Check your illuminator’s manual.)
The lamp has burned out.
Replace the lamp with the specified type.
(→Check your illuminator’s manual.)
The mercury lamp is not attached.
Attach a designated lamp.
(→Check your illuminator’s manual.)
The mercury lamp’s connector is not
connected to the illuminator.
Connect it to the illuminator.
(→Check your illuminator’s manual.)
The mercury lamp burns
out soon after it is turned
on.
The lamp type is incorrect.
The lamp is at end of its life.
Replace the lamp with the specified type.
(→Check your illuminator’s manual.)
Contact your nearest Nikon representative for
advice if a new lamp will burn out soon after
replacement.