Section
4-AA
501
CALIBRATION
PERFORMANCE CHECK PROCEDURE
Introduction
This procedure checks the Electrical Performance
Requirements as listed in the Specification section in this
manual. Perform the internal adjustment procedure if the
instrument fails
tomeet thesechecks. If recalibration does
not correct the discrepancy, circuit troubleshooting is
indicated. Also, use this procedure to determine
accep-
tibility of performance in an incoming inspection facility.
For convenience, many steps in this procedure check the
performance of this instrument at only one value in the
specified performance range. Any value within the
specified range, within appropriate limits, may be sub-
stituted. The performance check may be done at any
ambient temperature between C and
+50°
C.
Test Equipment Required
The test equipment listed in Table4-1, or equivalent, is
suggested to perform the performance check and the
adjustment procedure.
Exercise caution as dangerous voltages may be
encountered in some of the following steps.
Table
4-1
SUGGESTED TEST EQUIPMENT
Description
1
---
Requirements
I
i
1
Equipment
Minimum
Low distortion
sinewave oscillator
with
IM test signal
Performance Adjustment
Check
1
Procedure Recommended
<0.0008°/o THD 20 Hz
to 20 kHz;
<0.0018%,
10 Hz to 20 kHz and
20 kHz to 50 kHz;
<0.0032% 50 kHz to
100 kHz. 60
mV to 36 V
rms, 10 Hz to 100 kHz.
Sinewave oscillator Sinewave 600 /IV to 4, 6,
9,
10, 11,
1
9
1 1 A
I
TEKTRONIXSG 502
6,
7,
8,
9,
10, 11.
11A, 12, 13, 14
(2 required for 35
v
rms
alternate Step
11A) Frequency 20 Hz to
>300 kHz.
Ac voltage calibrator
1
100 ,uV to 180 V
8,
9,
10, 11, 11A
10 Hz to 100 kHz
I
TEKTRONIX SG 505
Fluke
5200A and
Dv m
(
lm~to2~ 12, 13, 14
I
1
1
TEKTRONIX DM 505
Counter
Bnc male to dual
60 Hz to 84 kHz
@
0.9 V
binding post adapter
I
I
I
I
12
1, 2,
5,
13, 14, 15
cia
4-
1
TEKTRONIX DC 504
Scans
by
ARTEK
MEDL4
=>