Performance tests
Marker high and low level accuracy
Required equ ipment Prerequisites
Digital multimeter
BNC-dual banana adapter
50 BNC cable
50 BNC termination
SMA male-BNC female adapter
50 SMA termination
(See page 2-21, Prerequisites.)
+ Marker high level
accuracy
1. Perform the Measure the termination resistance procedure. (See page 2-24,
Measure the termination resistance.)
2. Press the All Outputs On/Off button on the instrument to turn off all the
outputs.
3. Use the 50 SMA cable, SMA-BNC adapter, 50 BNC termination, and
BNC-Banana adapter to connect the CH 1 Markers M1 + connector on the
instrument to the HI and LO inputs on the digital multimeter.
4. Use the 50 SMA termination to terminate the CH 1 Markers M1
connector on the instrument.
Figure 2-17: Equipment connection for verifying the marker high and low level
accuracy