Specifications
Table 1-3: Samp
le clock generator
Characteristics Description
Instrument / Option Sampling rate control range
AWG70001A Option 150 1.49 kS/s to 50 GS/s
AWG70002A Option 225 1.49 kS/s to 25 GS/s
AWG70002A Option 216 1.49 kS/s to 16 GS/s
AWG70002A Option 208 1.49 kS/s to 8 GS/s
Resolutio
n
With jitte
r reduction: 3 digits
Without jitter reduction: 8 digits
Phase adj
ustment
Used to ma
nually align the outputs with other generators or instruments.
Range
-10,800
° to +10,800°
Resolution
1°
Table 1 -4: Analog output
Characteristics Description
Connector type Aeroflex/Weinschel Planar Crown Universal Connector System with SMA female adapter
Number of outputs AWG70001A: 1
AWG70002A: 2
Type of outputs (+) and (–) complementary output
ON/OFF control Independent control for each analog output complementary pair
Output impedance 50 Ω
VSWR
AWG70001A DC to 5 G Hz = 1.32:1
5 GHz to 10 GHz = 1.52:1
10 G Hz to 20 GHz = 1.73:1
AWG70002A DC to 5 G Hz = 1.61:1
5 GHz to 10 GHz = 1.61:1
Amplitude
Independent control for each channel
Range
0.25 V
p-p
to 0.5 V
p-p
into 50 Ω single-ended
0.5 V
p-p
to 1.0 V
p-p
into 100 Ω differential
Resolution 1 mV
DC accuracy (warranted)
±(2% of amplitude + 1 mV)
Amplitude flatness Flatness is measured as the deviation from the ideal sin(x)/x response curve of the DAC at the
specified sample rate. The response is mathematically removed from the measured data.
AWG70001A +1.8 dB to –1.8 d B up to 10 GHz
+1.8 dB to –3 dB from 10 GHz to 15 GHz
AWG70002A +0.8 dB to –1.5 d B up to 10 GHz
AWG70000A Series and AWGSYNC01 Technical Reference 1–3