Performance tests
Test record
Photocopy this page and the next two pages, and use them to record the
performance test results for your instrument.
AWG70000A series performance test record
Instrument Model:
Instrument Serial N umber: Certificate Number:
Temperature:
RH %:
Date of Calibration:
Technician:
Performance Test Minimum Incoming O utgo ing Maximu m
10 MHz R eference Frequency Accuracy
9.99999 MHz 10.00001 MHz
Analog Amplitude Accuracy
AWG70001A, AWG70002A
Amplit
ude
250 mV
p-p
244 mV 256 mV
375 mV
p-p
367 mV 383 mV
Ch1+
500 mV
p-p
489 mV 511 mV
Amplitude
250 mV
p-p
244 mV 256 mV
375 m
V
p-p
367 m
V
383 m
V
Ch1–
500 mV
p-p
489 mV 511 mV
AWG70002A
Amplitude
250
mV
p-p
244
mV
256
mV
375 mV
p-p
367 mV 383 mV
Ch2+
500 mV
p-p
489 mV 511 mV
Am
plitude
250 mV
p-p
244 mV 256 mV
375 mV
p-p
367 mV 383 mV
Ch2–
5
00 mV
p-p
4
89 mV
5
11 mV
Marker High and Low Level Accuracy
AWG70001A, AWG70002A
AWG70000A Series and AWGSYNC01 Technical Reference 2–35