To use.
â–
Click the Trigger Type box and select Bus from the drop-down list.
â–
Click the Bus box and select the bus number or name from the drop-down list.
NOTE. You have the option to add a user-defined label for bus sources.
â–
Click the Logic Thresholds Setup button to set the voltage threshold levels for the channels in the bus.
â–
Select the type of bus cycle or activity to use as the trigger from the Trigger On drop-down list.
â–
For some Trigger On selections, you must set additional fields to define other parameters such as for a Pattern.
In this example, you need to set the Character or Symbol values and select the Disparity. You can also set the component
threshold levels through the Logic Thresholds Setup button.
Click the Trigger On selections Character/Symbol or Error to view their parameters.
What do you want to do next?
Learn about bus setups.
Trigger on Character/Symbol
Trigger on Error
The error type options available in Trigger on Error are as follows:
Trigger setups
DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series 397