14-12
IM WT5000-01EN
Integration resume action at power failure recovery (Resume
Action)
You can set how to resume the integration operation when the instrument is turned off due to a power failure or
interruption and then turns back on.
Restart (Start)
The integration result at the time the power is turned off is stored. When the power recovers, integration starts
(continues) automatically.
Stop (Stop)
The integration result at the time the power is turned off is stored. When the power recovers, the integration
result up to the point when the power was turned off is displayed in the integration stopped state. You can start
(continue) integration.
Error (Error)
The integration result at the time the power is turned off is stored. When the power recovers, the integration
result up to the point when the power was turned off is displayed in the integration error state. You can start
integration by resetting it. When integration is reset, the measurement result is displayed as “-----” (no data). If
necessary, record the integration result up to that point before resetting it.
14 Integrated Power (Watt Hours/Ampere Hours)