8-12
IM WT5000-01EN
Master/Slave Synchronous Measurement (Sync Measure)
Connect the external start signal input/output terminals (MEAS. START) of the master and slave instruments
using a BNC cable (sold separately). You can synchronize the measurement of two instruments making one
the master and the other the slave. The master outputs a measurement start signal, and the slave receives the
signal.
For the specifications of the external start signal input/output terminals, see section 4.3 in the Getting Started
Guide, IM WT5000-03EN.
The measurement of the master and slave units cannot be synchronized under the following conditions:
When the data update interval differs between the master and slave.
During real-time integration mode or real-time storage mode.
Follow the procedure below to hold values during synchronized measurement.
To hold the display: Hold the display of the master first.
To unhold the display: Unhold the display of the slave first.
8 Computation Function

Table of Contents

Other manuals for YOKOGAWA WT5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA WT5000 and is the answer not in the manual?

YOKOGAWA WT5000 Specifications

General IconGeneral
BrandYOKOGAWA
ModelWT5000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals