EasyManuals Logo
Home>Agilent Technologies>Inverter>33220A

Agilent Technologies 33220A User Manual

Agilent Technologies 33220A
363 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #291 background imageLoading...
Page #291 background image
290
Chapter 5 Error Messages
Execution Errors
5
-221 Settings conflict;
trigger output connector used by trigger external
When the
external
trigger source is selected (
TRIG:SOUR EXT
command
),
the function generator automatically disables the “trigger out” signal.
The rear-
panel
Trig
connector cannot be used for both operations at the
same time
.
-221 Settings conflict;
frequency reduced for user function
For arbitrary waveforms, the output frequency is limited to 6 MHz.
When you change to the arbitrary waveform function (APPL:USER or
FUNC:USER
command) from a function that allows a higher frequency,
the function generator will automatically adjust the frequency to 6 MHz.
-221 Settings conflict;
frequency changed for pulse function
For pulse waveforms, the output frequency is limited to 5 MHz.
When you change to the pulse function (APPL:PULS or FUNC:PULS
command) from a function that allows a higher frequency, the function
generator will automatically adjust the frequency to 5 MHz.
-221 Settings conflict;
frequency reduced for ramp function
For ramp waveforms, the output frequency is limited to 200 kHz.
When you change to the ramp function (APPL:RAMP or FUNC:RAMP
command) from a function that allows a higher frequency, the function
generator will automatically adjust the frequency to 200 kHz.
-221 Settings conflict;
frequency made compatible with burst mode
For an internally-triggered burst, the output frequency is limited to a
minimum of 2.001 mHz. The function generator has adjusted the
frequency to be compatible with the current settings.
-221 Settings conflict;
burst turned off by selection of other mode or modulation
The function generator will allow only one modulation, sweep, or burst
mode to be enabled at the same time. When you enable a modulation,
sweep, or burst mode, all other modes are turned off.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33220A and is the answer not in the manual?

Agilent Technologies 33220A Specifications

General IconGeneral
BrandAgilent Technologies
Model33220A
CategoryInverter
LanguageEnglish

Related product manuals