The Logging Application Applications
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 251
Figure 153 The Agilent 8163B Logging Analysis Screen - First Screen
Setting Analysis Parameters
3 To edit the power units used for the analysis, move to [W], press Enter, move to
<dBm>, <W>, or <dB>, and press Enter. See What are the Power Units ?” on
page 85 for an explanation of power units.
4 To edit the reference mode used for the analysis, move to <Ref Mode>, press
Enter, move to one of the following referencing modes:
•<Value>, you set a reference value using <Ref>, see step 5,
•<First Sample>, the first sample of the logging function is selected as the refer-
ence value,
•<Channel 1>, the value that Channel 1 measures is updated as the reference
value for each set of power measurements (thus, each data value for channel 1
will be 0 dB), and
•press Enter.
5 If you have chosen <Value> as the <Ref Mode>, move to <Ref>, press Enter,
enter a value for the reference, and press Enter.
These settings have an effect when you choose <dB> as the <Pwr unit>.
See step 8 or step 3.
NOTE

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 8163A and is the answer not in the manual?

Agilent Technologies Agilent 8163A Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 8163A
CategoryMultimeter
LanguageEnglish

Related product manuals