Error Messages Attenuator Modules
414 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
Attenuator Modules
The following Error Codes are applicable to attenuator modules.
Codes E10010, E10012, and E10014 are valid for all attenuators. The remaining
codes are only valid for attenuators with power control.
error message E10010: Invalid Caldata
description The Caldata loaded on the module is invalid. Either the caldata ver-
sion number is not compatible with the firmware or the caldata is
corrupt (checksum incorrect)
test value (no test value)
action Please return the unit to your local service center for repair
error message E10012: Dual ported RAM write/read test failed
description Writing bit patterns and reading them back to/from the dual ported
ram failed
test value (no test value)
action Please return the unit to your local service center for repair
error message E10013: FPGA not correctly initialized
description Loading the FPGA with data failed. This indicates a defective FPGA
or an FPGA access problem
test value (no test value)
action Please return the unit to your local service center for repair
error message E10014: Firmware corrupt, checksum error (a,b)
description Firmware in flash has an invalid checksum. Either the image which
was programmed was corrupt or the flash memory has a failure
test value a is the stored checksum
b is the calculated checksum
action Reload the module firmware. If the failure persists, please return the
unit to your local service center for repair
error message E10031: 0.25 ref out of limits (a, b, c)
description The 0.25V reference voltage measured through the 16bit ADC is out
of the acceptable limits
test value a shows the ADC reading (0.25V = 3276)
b shows the lower limit
c shows the upper limit
action Please return the unit to your local service center for repair