Equipment Required Performance Tests
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 353
Test Record
Results of the Performance Test may be tabulated in the Test Record
provided after the test procedures. It is recommended that you fill out the
Test Record and refer to it while doing the test. Since the test limits and
set-up information are printed on the Test Record for easy reference, the
record can also be used as an abbreviated test procedure (if you are
familiar with test procedures). The Test Record can also be used as a
permanent record and may be reproduced without written permission from
Agilent Technologies.
Test Failure
If the Agilent 8163A/B Lightwave Multimeter, the Agilent 8164A/B
Lightwave Measurement System, or the Agilent 8166A/B Lightwave
Multichannel System fail any Performance Test, return the instrument to
the nearest Agilent Technologies Sales/Service Office for repair.
Instruments Specifications
Specifications are the characteristics of the instrument that are certified.
These specifications, listed in Appendix C are the limits against which the
Agilent 8163A/B Lightwave Multimeter, the Agilent 8164A/B Lightwave
Measurement System, or the Agilent 8166A/B Lightwave Multichannel
System can be tested. Appendix C also lists some supplemental
characteristics of the Agilent 8163A/B Lightwave Multimeter, the
Agilent 8164A/B Lightwave Measurement System, and the
Agilent 8166A/B Lightwave Multichannel System and should be
considered as additional information.
Any changes in the specifications due to manufacturing changes, design,
or traceability to the National Institute of Standards and Technology will be
covered in a manual change supplement or revised manual. The
specifications listed here supersede any previously published.
Make sure that all optical connections of the test setups given in the
procedure are dry and clean. DO NOT USE INDEX MATCHING OIL. For
cleaning, use the cleaning instructions given in Appendix E.
NOTE

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 8163A and is the answer not in the manual?

Agilent Technologies Agilent 8163A Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 8163A
CategoryMultimeter
LanguageEnglish

Related product manuals