Agilent 8166A Specifications Specifications and Regulations Compliance
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 341
Agilent 8166A Specifications
The Agilent 8166A Lightwave Multichannel System displays 17 channels.
[1]
Modules designed for the 8153A Lightwave Multimeter are NOT
compatible with the Agilent 8166A Lightwave Multichannel System.
[2]
For code compatibility to the 8153A Lightwave Multimeter please refer
to the mainframe’s Programming Guide, part number 08164-91016.
Display Graphical display 190 × 300 points visible, monochrome.
Display Resolution (Power Meters) 0.0001 dB/dBm, 0.01 pW to 10 pW (depending on power
range), up to 6.5 digits (user definable).
Compatibility The Agilent 8166A supports all Agilent 8163A Series
modules.
[1]
Trigger True synchronous on all channels
Environmental
Storage temperature
Operating temperature
Humidity
-40° C to +70° C
0° C to +45° C
<95% R.H. from 0° C to +45° C
Power AC 100 - 240 V ± 10%, 50 - 60 Hz, 450 VA max.
GPIB interface
Function code
All modes and parameters accessible via GPIB interface
SCPI standard
[2]
Interfaces Parallel port (Centronics)
Serial port (RS232)
Number of Modules 17 Modules
Dimensions 240 mm × 430 mm × 580mm (9.5" × 17" × 23")
Weight net 23 kg (50.7 lbs); shipping 45 kg (94.8 lbs)