How to Use Auxiliary Functions Tunable Lasers
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 155
How to Perform a Lambda Zero
Performing a Lambda Zero recalibrates the optical wavelength. This
wavelength may drift due to a change in temperature and other
environmental conditions.
A Lambda Zero is automatically performed when the instrument boots or
when an Automatic Realignment is performed.
The instrument automatically performs a wavelength zero from time to
time. To avoid interruption of your measurements, you should perform a
wavelength zero at a time of your choice.
To perform a wavelength zero:
1 Move to the Tunable Laser channel and press Menu.
2 Move to <λ Zeroing> and press [OK]. You will see the screen in Figure 82 .
Figure 82 λ Zeroing Screen
3 Wait several minutes. This time depends on how much the wavelength of the
instrument has drifted since the last Lambda Zero was performed.
The instrument automatically performs a Lambda Zero if there is a ± 4.4 K
temperature change since the last Lambda Zero was performed.
NOTE

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 8163A and is the answer not in the manual?

Agilent Technologies Agilent 8163A Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 8163A
CategoryMultimeter
LanguageEnglish

Related product manuals