Installation and Maintenance GPIB Interface
312 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
GPIB Interface
You can connect your GPIB interface into a star network, a linear network,
or a combination star and linear network. The limitations imposed on this
network are as follows:
The total cable length cannot exceed 20 meters
The maximum cable length per device is 2 meters
No more than 15 devices may be interconnected on one bus.
Cables and Adapters
See GPIB Cables and Adapters” on page 333 for details on cables and
adapters.