Optical Switch modules Error Messages
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 417
Optical Switch modules
error message InitError() failed!
description Initialization of internal error buffer failed.
test value
action Please return the unit to your local service center for repair
error message Invalid Caldata!
description The calibration data loaded on the module is invalid.
Either the version number is nir compatible with the firmware or the
calibration data is corrupt (checksum incorrect).
test value
action Please return the unit to your local service center for repair
error message InitModule() failed!
description Module initialization failed.
note Normally something more descriptive appears (i.e. one of the other
messages mentioned here.)
test value
action Please return the unit to your local service center for repair
error message SanityChecks():channel X, configArray != -1
SanityChecks():channel X, configs == 0
SanityChecks() failed
description Calibration data seems not to be correct
test value X Channel number
action Please return the unit to your local service center for repair
error message InitModuleStruct() failed
description Initialization of module structures failed
test value
action Please return the unit to your local service center for repair
error message Unsupported module type
description Module type described in calibration data not supported by the firm-
ware or incorrect calibration data
test value
action Please return the unit to your local service center for repair