Using Assist/Control
In Assist/Control mode (A/C), the patient may trigger the
controlled ventilator breaths within a time period (trigger
window). Ventilator breaths will then be synchronized
with the patient's inspiratory effort!
NOTE: In this mode, the resulting ventilation rate is higher
than that set on the ventilator.
Ventilator breaths are automatically applied at the set rate
if there is no patient inspiratory effort.
Set ventilation initially as described for CMV mode,
pages 29, 30.
Then:
1 Press »
M
Info« key to obtain the display:
CMV A/C.
2 Press »g Reset« to confirm and Assist/Control mode
is applied by the ventilator.
The maximum inspiratory time is always limited to 1.3
seconds. The following alarm message appears if the
measured rate exceeds the set frequency by more than
50 %:
High rate
Operation
Using Assist/Control
Operating Instructions Oxylog 2000, SW 3.n
32
Oxylog 2000
D
12
00729033

Table of Contents

Other manuals for Dräger Oxylog 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 and is the answer not in the manual?

Dräger Oxylog 2000 Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals