Technical Data
Ambient Conditions
For operation
Temperature - 18 to 50°C (0 to 122ºF)
Atmospheric pressure 600 to 1200 hPa
Relative humidity 30 to 95% relative humidity
For storage
Temperature – 18 to 70°C (0 to 158ºF)
Atmospheric pressure 600 to 1200 hPa
Relative humidity 10 to 95% relative humidity
Performance Data
Operating parameters
Control principle Volume flow controlled, time cycled, volume constant flow chopper
(microprocessor controlled)
spontaneous breathing with integrated demand valve
(also at elevated PEEP levels)
Ventilation modes CMV / A/C, SIMV/CPAP
Ventilator rate 5 to 40 bpm ± 1 bpm
infinitely adjustable
Tidal volume VT 0.1 to 1.5 L
infinitely adjustable
Setting tolerance at ± 10% of set value,
10 cmH2O airway pressure 50 mL minimum
I:E ratio 1:3 to 2:1 ± 5%
infinitely adjustable
Inspiratory pressure 20 to 60 cmH2O ±10%
limitation Pmax infinitely adjustable;
can also be set to 80 cmH2O ±10%
minimum ± 3 cmH2O
PEEP (0 +2) to 15 cmH2O,
infinitely variable
Expiratory minute min. 1.0 L/min
ventilation MV max. 25.0 L/min
with I:E = 1:1.5
flow range for these
MV-values 4 to 60 L/min
Technical Data
Ambient Conditions
Performance Data
Operating Instructions Oxylog 2000, SW 3.n
51

Table of Contents

Other manuals for Dräger Oxylog 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 and is the answer not in the manual?

Dräger Oxylog 2000 Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals