Maintenance
Preventive Maintenance Intervals
User Replaceable Parts
Operating Instructions Oxylog 2000, SW 3.n
43
User Replaceable Parts
Batteries
WARNING !
Treatment of batteries:
- Do not throw into fire!
- Do not force open!
- Do not attempt to recharge
alkaline batteries!
Danger of bodily injury.
Follow all local, state, and federal regulations
with respect to environmental protection when
disposing of batteries.
Internal alkaline battery
pack
replace, if message
»Bat. discharged« or
»Change bat.«
is displayed
Internal NiCd battery pack replace, if message
»defective NiCd« is
displayed during charging,
every 2 years at the latest
Maintenance
Preventive Maintenance Intervals
Scheduled safety checks every 2 years
Inspection and prev. maint.
every 2 years
Pressure regulator overhaul
every 6 years
WARNING !
Preventive Maintenance work on the Oxylog
2000 ventilator may be performed by factory
trained and authorized personnel only.
WARNING !
To avoid any risk of infection, clean and disin-
fect ventilator and accessories before any
maintenance according to established hospital
procedures - this applies also when returning
ventilators or parts for repair.
CAUTION !
Maintenance
In case of malfunction of this device, contact
your local Dräger Service or our Factory
Authorized Technical Service Center.
The devices must be inspected and serviced
(preventive maintenance) by factory trained and
authorized technical service representatives at
regular intervals. A record must be kept on this
preventive maintenance. We recommend obtain-
ing a service contract through your vendor.
Maintenance or repair of Oxylog ventilators shall
be performed only by Dräger authorized techni-
cal service representatives.