Operation
Using SIMV
Operating Instructions Oxylog 2000, SW 3.n
33
Using SIMV
SIMV = Synchronized IntermittentMmandatory Ventilation
SIMV is a combination of mechanical ventilation and
spontaneous breathing, with spontaneous breathing
possible between mandatory ventilator breaths.
If the patient does not breathe spontaneously within a
certain time period, the ventilator automatically applies a
mechanical breath.
A 6 second trigger window is kept open when setting
low ventilation rates. This will guarantee a minimum
expiratory time of 0.5 seconds. The duration of the trigger
window decreases as the frequency of mechanical
breaths increases.
With synchronization, mechanical breaths are applied
during a trigger time window that is calculated to
coincide with the patient's inspiratory effort. The
frequency of the mechanical breaths remains constant.
Synchronization is briefly indicated by a star (*) in the
display.
For rates of 5 to 40 bpm:
The mechanical breaths and their time parameters are
determined by the ventilator rate set for SIMV/CPAP:
»Rate« setting Effect
0 No mechanical breaths CPAP
(page 35)
5 to 18.5 bpm Fixed inspiratory time = 1.3 s
18.5 to 40 bpm Fixed I : E = 1 : 1.5

Table of Contents

Other manuals for Dräger Oxylog 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 and is the answer not in the manual?

Dräger Oxylog 2000 Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals