Instructions for Use Oxylog 3000 SW 1.n 101
Cleaning, Disinfection and Sterilization
The disassembled parts of the breathing valve,
the flow sensor, the angled connector, the flow
measuring hoses and the ventilation hose can
be sterilized in hot steam at 134
o
C in accor-
dance with EN 285 (Sterilization – Steam
sterilization – Large-scale sterilization) for at
least 3 minutes, up to 10 minutes.
The hose system can be sterilized up to a
100 times.
Sterilization for longer than 10 minutes is
permissible, but will decrease the service life of
the hose system.
After care
Reassemble, refer to "Assembly" on page 27
for information.
Connect to the power supply and gas supply,
refer to "Assembly" on page 27 for information.
Check readiness for operation, refer to"Getting
Started" on page 37 for information.
Note service life of the hose system
The parts of the breathing valve, the flow sensor,
the angled connector, the flow measuring hoses
and the ventilation hose are resistant to the recom-
mended disinfectants and to the temperatures
occurring during sterilization.
However, every disinfection and sterilization cycle
also means wear on the parts concerned.
For this reason, the parts must be examined for
cracks and permanent deformation after the
cleaning procedure.
NOTE
Damaged or deformed parts must be replaced.