Assembly
30
Instructions for Use Oxylog 3000 SW 1.n
When using a bacterial filter or HME
z Connect the bacterial filter or HME to the angled
connector.
Hose connections
1 Connect the ventilation hose (A) to the
breathing valve.
2 Connect the flow measuring hoses (B) to the
nozzles on the flow sensor. Note the different
diameters.
3 Connect the flow measuring (A) hoses to the
Oxylog 3000.
4 Connect the ventilation hose (B) to the gas
outlet on the Oxylog 3000.
NOTE
When using a bacterial filter or HME, measured
flows may deviate from the expiratory flows, as
temperature and humidity of the gas are reduced.
010
WARNING
HME and bacterial filters increase the
exhalation resistance and dead space volume
of the ventilation system.
011
A
B
B
WARNING
Do not use electrically conductive hoses!
Risk of electric shock.
This can endanger the patient.
012
B
A