EasyManuals Logo
Home>Dräger>Medical Equipment>Oxylog 3000

Dräger Oxylog 3000 User Manual

Dräger Oxylog 3000
138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #114 background imageLoading...
Page #114 background image
Technical Data
114
Instructions for Use Oxylog 3000 SW 1.n
Performance data
Control principle Time-cycled, volume-constant, pressure-controlled.
Maximum inspiratory flow 100 L/min
1)
Device compliance
with 1.5 m ventilation
hose
< 1 mL/mbar
with 3 m ventilation hose <
2 mL/mbar
Inspiration resistance <
6 mbar at 60 L/min
<
4 mbar at 30 L/min
< 2 mbar at 5 L/min
Expiration resistance <
6 mbar at 60 L/min
<
4 mbar at 30 L/min
< 2 mbar at 5 L/min
Dead space including flow sensor approximately 35 mL (reusable hose system).
approximately 33 mL (disposable hose system).
Supplementary functions
Demand valve Opens the breathing system upon failure of the gas
supply, permits spontaneous breathing with ambient
air.
Relief valve Opens the breathing system at approximately
80 mbar.
Patient connection 22 mm ISO conical connector.
1) At service pressures >350 kPa.
The maximum inspiratory flow is reduced to 80 L/min at service pressures <350 kPa and to 39 L/min at service
pressures <280 kPa.