EasyManuals Logo
Home>Dräger>Medical Equipment>Oxylog 3000

Dräger Oxylog 3000 User Manual

Dräger Oxylog 3000
138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background imageLoading...
Page #39 background image
Instructions for Use Oxylog 3000 SW 1.n 39
Getting Started
Determining the approximate pneumatic operating time for the
Oxylog 3000
Example for supply of medical gas:
Cylinder pressure measured on the pressure
gauge of the pressure reducer: 2000 kPa
Liquid capacity of the O
2 cylinder: 2.1 L
Supply of medical gas:
2.1 L x 2000 kPa = approximately 420 L
Example for pneumatic operation time:
IPPV mode, frequency 10 breaths /min,
VT = 1 L, O
2 = 100%
Minute volume = 10 breaths /min x 1 L =
10 L/min
Operation time =
* Calculated with average gas consumption of
ventilator: 0.5 L/min
Operation time = = approx. 40 minutes
The pneumatic operation time increases when
Oxylog 3000 operates with O
2 concentration of less
than 100 % O
2, as ambient air is drawn into the
device.
The amount of gas from the high-pressure supply,
which is currently being consumed, is indicated by
the Oxylog 3000 in the lower left section of the
information window in L/min. This display is over-
written when a higher priority message is activated.
Example:
A O
2 consumption = 2.5 L/min
Medical gas supply [L]
(MV +0.5*) [L/min]
420
10.5
13637171
A

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 3000 and is the answer not in the manual?

Dräger Oxylog 3000 Specifications

General IconGeneral
BrandDräger
ModelOxylog 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals