Instructions for Use Oxylog 3000 SW 1.n 81
Configuration
Set startup settings
Range of the settings:
The default settings for the parameters are
displayed on the screen when the ventilator is
switched ON. The settings can be adjusted.
Advance to the second page:
1 Select the line Page, confirm and turn rotary
knob.
To restore the manufacturer's defaults:
2 Select and confirm line Set factory default.
Select hose type
The type of ventilation hose (reusable or
disposable hose system) can be configured.
The gas outlet connectors for the flow measuring
hoses must correspond to the hose type system
selected.
Parameter Range
Trigger 0 (OFF) 3 to 15 L/min
PEEP 0 to 20 mbar
I:E 3:1 to 1:4
Tinsp 0.2 to 10.0 s
Tplat % 0 to 50%
ΔASB 0 to 35 mbar
Ramp SLOW, STANDARD, FAST
Pinsp 3 to 55 mbar
O
2-Flow 0 to 15 L/min
NIV ON, OFF
Tapn 0 (OFF) to 60 s
VTapn 50 to 2000 mL
fapn 12 to 60 bpm
MV-high 2.0 to 41 L/min
MV-low 0.5 to 40 L/min
fspn-high 10 to 100 bpm
Loudness 1/4 to 4/4
Brightness 1/4 to 4/4