EasyManuals Logo
Home>Dräger>Medical Equipment>Oxylog 3000

Dräger Oxylog 3000 User Manual

Dräger Oxylog 3000
138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
Instructions for Use Oxylog 3000 SW 1.n 21
Operating Concept
Ventilation controls
A Keys for selecting the ventilation modes:
IPPV (CMV), SIMV, CPAP
BIPAP
®
/ASB (PCV+/PS)
SIMV/ASB (SIMV/PS)
CPAP/ASB (CPAP/PS).
B Ventilation parameter controls
– Inspiratory tidal volume VT [mL],
– Ventilation frequency Freq. [1/min],
– Maximum inspiratory pressure Pmax [mbar],
– O
2 Concentration O2 [%].
C Rotary knob.
For primary care Oxylog 3000 starts in IPPV venti-
lation mode with user -configured starting values
for I:E (1.5 as default setting) and PEEP (5mbar as
default setting).
Selecting the ventilation mode
z Press the appropriate ventilation mode key (A)
for approximately 3 seconds.
Or
1 Press the appropriate ventilation mode key (A).
2 Press the rotary knob (C) to confirm. The
selected ventilation mode will be activated.
The active ventilation mode is displayed in the
upper left corner of the display (B).
Refer to the section "Operation" on page 45 for
additional information on ventilation mode setting.
00137170
NOTE
Different ventilation modes and their parameters
can be set in the display window via the rotary
knob (e.g. Tinsp, PEEP, Δ ASB, Pinsp).
To select the parameter: turn rotary knob.
To activate the parameter: press rotary knob.
To set the value: turn rotary knob.
To confirm the value: press rotary knob.
CB
A
08537170
Oxylog 3000
CB
A

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 3000 and is the answer not in the manual?

Dräger Oxylog 3000 Specifications

General IconGeneral
BrandDräger
ModelOxylog 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals