Instructions for Use Oxylog 3000 SW 1.n 35
Assembly
Oxylog 3000 can be supplied with either O2 or
medical air.
z Set the gas type in the configuration menu to
ensure correct metering, "Check connections"
on page 42.
Supply from an O2 cylinder
1 Use a full O2 cylinder.
2 Connect the pressure reducer (270 to 600 kPa
delivery pressure, 500 kPa nominal pressure)
to the O
2 cylinder.
3 Connect the O2 medical gas hose (A) to the
Oxylog 3000.
4 Connect the O
2 medical gas hose to the
pressure reducer (B).
WARNING
No smoking or open flames.
O
2 enhances combustion of other substances
and can intensify fires.
WARNING
Only use compressed gas cylinders and pres-
sure reducers, which comply with all applica-
ble regulations and have been approved.
WARNING
Only use a pressure reducer with a relief valve
at the outlet to limit the delivery pressure to a
maximum of 1000 kPa in case of a malfunc-
tion, to prevent damage to the ventilator!