Monitoring
72
Instructions for Use Oxylog 3000 SW 1.n
O2 inhalation (optional)
If the gas flow is obstructed, the flow is interrupted
by the ventilator for 5 sec at an airway pressure of
30 mbar and the airway pressure is reduced to
0mbar.
The !!! Paw high alarm is active.
z The spontaneously breathing patient may only
be connected to the device via an inhalation
mask.
To activate O
2 inhalation
1 Press and hold key O
2-Inhalat. (A) for approx.
3 seconds
or
2 Briefly press key O2-Inhalat. (A) and confirm
with rotary knob (C).
3 Connect the inhalation mask to the inspiration
connector (B).
WARNING
The O2 inhalation function is not a ventilation
mode!
It may only be used for patients with spontane-
ous breathing who receive a constant O
2 flow
of between 0 and 15 L/min via a mask.
08337170
Oxylog 3000
C
A B