EasyManuals Logo
Home>Dräger>Medical Equipment>Oxylog 3000

Dräger Oxylog 3000 User Manual

Dräger Oxylog 3000
138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #50 background imageLoading...
Page #50 background image
Operation
50
Instructions for Use Oxylog 3000 SW 1.n
Cardio-pulmonary Resuscitation (CPR)
During CPR, the airway pressure Paw is increased
because of chest compressions.
The Oxylog 3000 will try to limit the airway pressure
Paw to the set Pmax, without ending the inspiration
prematurely.
However, if due to compressions the airway
pressure Paw exceeds the set Pmax by 5 mbar,
the Oxylog 3000 cycles to the expiration phase.
Therefore in general, if Pmax is set to a higher
value, a higher minute volume is possible.
However, this increases the intra-thoracic pressure
and may reduce coronary perfusion.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 3000 and is the answer not in the manual?

Dräger Oxylog 3000 Specifications

General IconGeneral
BrandDräger
ModelOxylog 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals