Emerson Process Management GmbH & Co. OHG A-17
A
Appendix
X-STREAM XE
Short Form Manual
HASXEE-SFM-HS
05/2017
A.2 Assignment of Terminals and Sockets
A.2.1 Tabletop & Rack Mount Analyzers
A.2 Assignment of Terminals and Sockets
DC 24 V Input (
1
2 19" analyzer)
Socket X1 - Analog outputs, relay outputs 1...5
(Assignment of screw terminals adaptor: see XSTA on next page)
Socket X4 - Digital I/O
(Assignment of screw terminals adaptor: see XSTD on next page)
Pin 1
Pin 8
Connector X2 - IOIOI - Serial Modbus interface
(Assignment of screw terminals adaptor: see XSTA on next
page)
Ethernet connector
for Modbus
Connector X5 - Analog inputs
(Assignment of screw terminals adaptor: see XSTI on next page)
Pin no.
Signal
1 TX+
2 TX-
3 RX+
6 RX-
other not used
Note!
Conguration of relay contacts as per standard
factory setting (NAMUR status signals). NAMUR
status signals are automatically congured Fail
Safe!
Signal Pin
Analog
outputs
Channel 1, (+) 4 (0) - 20 mA 1
Channel 2, (+) 4 (0) - 20 mA 2
Channel 3, (+) 4 (0) - 20 mA 3
Channel 4, (+) 4 (0) - 20 mA 4
Channel 5, (+) 4 (0) - 20 mA 5
unused 6
unused 7
Relay contacts
(Status signals)
Output1 (Failure) NC 8
Output1 (Failure) NO 9
Output1 (Failure) COM 10
Output3 (Off spec) NC 11
Output3 (Off spec) NO 12
Output3 (Off spec) COM 13
Pin Signal
14 Channel 1, GND
Analog
outputs
15 Channel 2, GND
16 Channel 3, GND
17 Channel 4, GND
18 Channel 5, GND
19 unused
20 Output2 (Maintenance request) NC
Relay contacts
(Status signals)
21 Output2 (Maintenance request) NO
22 Output2 (Maintenance request) COM
23 Output4 (Function check) NC
24 Output4 (Function check) NO
25 Output4 (Function check) COM
Signal Pin
Input 1 1
Input 2 2
Input 3 3
Input 4 4
GND for all digital inputs 5
unused 6
unused 7
Output 5 NC 8
Output 5 NO 9
Output 5 COM 10
Output 6, NC 11
Output 6, NO 12
Output 6, COM 13
Output 7, NC 14
Output 7, NO 15
Output 7, COM 16
Output 8, NC 17
Output 8, NO 18
Output 8, COM 19
Digital
outputs
Digital
inputs
Pin Signal
20 Input 5
21 Input 6
22 Input 7
23 Output 9, NC
24 Output 9, NO
25 Output 9, COM
26 Output 10, NC
27 Output 10, NO
28 Output 10, COM
29 Output 11, NC
30 Output 11, NO
31 Output 11, COM
32 Output 12, NC
33 Output 12, NO
34 Output 12, COM
35 Output 13, NC
36 Output 13, NO
37 Output 13, COM
Digital
outputs
Digital
inputs
Note!
The conguration illustrated
here is that of the rst socket,
labelled X4.1.
Inputs 8-14 and outputs 14-
22, are on the second socket
(X4.2), if installed.
Pin Signal
9 Input 2 high (+)
8 Input 2 low (-)
7 Input 1 low (-)
6 Input 1 high (+)
Signal Pin
BR 2 t a wire bridge here to apply analog
5
BR 2 signal in current mode to input 2
*)
4
unused 3
BR 1 t a wire bridge here to apply analog
2
BR 1 signal in current mode to input 1
**)
1
{ }
{ }
*)
alternatively set jumper P1 on electronics board XASI
**)
alternatively set jumper P2 on electronics board XASI
1: ME
2: + 24 V
3: 0 V ()
2
1
3
Pin
6
7
8
9
Pin
1
2
3
4
5
Pin no.
MOD 485/
2 wire
MOD 485/
4 wire
RS 232
1 Common Common Common
2
not used not used
RXD
3
not used not used
TXD
4
not used
RXD1(+)
not used
5 D1(+) TXD1(+) Common
6
not used not used not used
7
not used not used not used
8
not used
RXD0(-)
not used
9 D0(-) TXD0(-)
not used

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount Analytical X-STREAM Enhanced Series and is the answer not in the manual?

Emerson Rosemount Analytical X-STREAM Enhanced Series Specifications

General IconGeneral
BrandEmerson
ModelRosemount Analytical X-STREAM Enhanced Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals