Emerson Process Management GmbH & Co. OHG 1-1
1
Techn. Description
X-STREAM XE
Short Form Manual
HASXEE-SFM-HS
05/2017
Chapter 1
Technical Description
The following are the main features of the
Emerson Process Management X-STREAM
Enhanced (hereinafter also referred to as
"X-STREAM XE") gas analyzers in brief:
compact design with easily accessible
internal components
customizable for a wide range of appli-
cations: different housings are available
while internal construction remains largely
identical
a highly integrated mainboard contains
all interfaces and basic functions for the
operation of the unit
multilingual microprocessor-controlled
user interface with liquid crystal (LCD) to
indicate measurement value and status
messages
units for outdoor use are optionally sup-
plied with an impact tested front panel
widerange power supply unit for worldwide
use without modication (
1
2 19in units with
internal or external PSUs)
X-STREAM Enhanced gas analyzers can
measure up to ve different gas components
using many combinations of the following
analyzing techniques:
IR = non-dispersive infrared analysis
UV = ultraviolet analysis
pO
2
=
paramagnetic oxygen analysis
eO
2
=
electrochemical oxygen analysis
tO
2
= trace oxygen analysis
TCD = thermal conductivity analysis
tH
2
O = trace moisture measurement
Modied resistant measuring cells are avail-
able for use with corrosive gases and/or
gases containing solvents.
Special congurations for the analysis of am-
mable gases are also available.
EXPLOSION HAZARD
X-STREAM XE analyzers MUST NOT be used in explosive environments
(hazardous areas) without additional safety features.
This instruction manual does NOT describe the special conditions necessary
to operate gas analyzers in hazardous areas.
Please refer to the separate instruction manual, supplied with units for use
in hazardous areas.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount Analytical X-STREAM Enhanced Series and is the answer not in the manual?

Emerson Rosemount Analytical X-STREAM Enhanced Series Specifications

General IconGeneral
BrandEmerson
ModelRosemount Analytical X-STREAM Enhanced Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals