10-5-2-2 Peripheral/Option Checks
If any peripherals or options are not part of the system configuration, the check can be omitted. Refer
to the User Manual for a list of approved peripherals/options.
10-5-3 Input Power
10-5-3-1 Mains Cable Inspection
10-5-4 Cleaning
10-5-4-1 General Cleaning
Table 10-39 GE Approved Peripheral/Hardware Option Functional Checks
Step Item Description
1VCR
Verify record/playback capabilities of the VCR. Clean heads and covers if necessary.
2 VGP Printer
Verify hardcopy output of the video graphic printer. Clean heads and covers if necessary.
3 B/W Printer
Verify hardcopy output of the B/W printer. Clean heads and covers if necessary.
4 Footswitch
Verify that the footswitch is functioning as programed. Clean as necessary.
Table 10-40 Mains Cable Inspection
Step Item Description
1Unplug Cord
Disconnect the mains cable from the wall and system.
2Inspect
Inspect it and its connectors for damage of any kinds.
3Verify
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that
no strands may cause a short circuit.
4 Verify
Inlet connector retainer is functional.
Table 10-41 General Cleaning
Step Item Description
1 Console
Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system. Be careful
not to get the cloth too wet so that moisture does not enter the console.
2 Probe Holder
Clean probe holders. (they may need to be soaked to remove excess gel).