GE MEDICAL SYSTEMS
DIRECTION 2317229, REVISION 3 LOGIQ™ 180 BASIC SERVICE MANUAL
10 - 16 Section 10-7 - Electrical Safety Tests
10-7-5-3 Data Sheet for Chassis Source Leakage Current
The test passes when all readings measure less than the value shown in Table 10-44. Record all data
on the PM Inspection Certificate.
Table 10-48 Typical Data Sheet for Chassis Source Leakage Current
Unit Power
Tester Polarity
Switch
Tester Neutral
or Ground
Switch
Test 1 Probe
Connector
Screw
Test 2
Rear
Panel
Optional
Test 4
Optional
Test 5
Enter Name of tested peripheral here:
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
OFF NORM OPEN
OFF NORM CLOSED
OFF REV OPEN
OFF REV CLOSED