GE HEALTHCARE PROPRIETARY TO GE
D
IRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 3-9 - Optional Peripherals/Peripheral Connection 3 - 37
Section 3-9
Optional Peripherals/Peripheral Connection
NOTE: For more detailed Safety Considerations when connecting auxiliary devices to the LOGIQ S8 system,
please review: Section 1-4-6 "Auxiliary Devices Safety" on page 1-20.
NOTE: Normally auxiliary devices and peripherals come pre-installed with the system.
WARNINGWARNING
!! WARNING:
After each installation, the leakage currents have to be measured according to
IEC 60601-1 respectively UL 60601-1.
CAUTION
!! CAUTION:
Please observe that some printers may not be medical devices! If the Line Printers are not
medical devices, they have to be located outside of the patient environment (according to IEC
60601-1 / UL 60601-1).