GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 2-3 - Facility needs 2 - 9
2-3-4 Desirable features
The followings are (optional) site recommendations.
Door opening of at least 76 cm (2.5 ft.) in width.
Accessible circuit breaker for a dedicated power outlet.
Sink with hot and cold running water.
Receptacle for bio–hazardous waste, for example, used probe sheaths.
Emergency oxygen supply.
Storage area for linens and equipment.
Nearby waiting room, lavatory, and dressing room.
Dual level lighting (bright and dim).
Lockable cabinet for software and manuals.
2-3-5 Minimal floor plan suggestions
Figure 2-1 Minimal Floor Plan 4.3m x 5.2m (14ft x 17ft)
Dedicated Power Outlets
Hospital Network
Cabinet for
Software and Manuals
(optional)
LOGIQ S8