EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ E8

GE LOGIQ E8 User Manual

GE LOGIQ E8
418 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #341 background image
GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 7-7 - Troubleshooting Trees, Instructions and Tech Tips 7 - 71
7-7-5 Trackball - Impaired sensitivity
Figure 7-64 Trackball - Impaired sensitivity
Trackball has low sensitivity.
Set up the appropriate trackball response time.
(via [Speed] and [Acceleration] in the Utility - System - General
- Trackball. Refer to Figure 3-23 on page 3-33)
Is the trackball issue resolved?
Is the trackball issue resolved?
Clean the trackball.
Refer to section 7-7-5-1 on page 7-72
Replace the trackball.
Return to normal scanning. Return to normal scanning.
No
No
No
Yes
Yes

Table of Contents

Other manuals for GE LOGIQ E8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E8 and is the answer not in the manual?

GE LOGIQ E8 Specifications

General IconGeneral
BrandGE
ModelLOGIQ E8
CategoryMedical Equipment
LanguageEnglish

Related product manuals