GE HEALTHCARE PROPRIETARY TO GE
D
IRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 3-7 - Power On/Off 3 - 17
Section 3-7
Power On/Off
NOTE: If the battery option is installed (R4 or later)
The extended battery/small battery is charged automatically when the LOGIQ S8 is connected to the
wall outlet via the power cable and circuit breaker is turned on whether the LOGIQ S8 is turned on or in
standby-mode.
But you can turn off the circuit breaker and disconnect the power cable when the battery is charging.
While the extended battery/small battery charging, a fan is running. a fan stops automatically when
battery is fully charged.
While the extended battery charging, the LED is blinking and turn off when the battery is fully
charged.
DANGER
ALWAYS CONNECT THE UNIT TO A FIXED POWER SOCKET WHICH HAS THE
PROTECTIVE GROUNDING CONNECTOR.
DANGER
NEVER USE A THREE-TO-TWO PRONG ADAPTER; THIS DEFEATS THE
SAFETY GROUND.
DANGER
ENSURE THAT THE POWER CORD AND PLUG ARE INTACT AND THAT THE
POWER PLUG IS THE PROPER HOSPITAL-GRADE TYPE (WHERE REQUIRED).
CAUTION
LOGIQ S8 REQUIRES ALL COVERS.
OPERATE THIS UNIT ONLY WHEN ALL BOARD COVERS AND FLAME PANELS ARE SECURELY
IN PLACE. THE COVERS ARE REQUIRED FOR SAFE OPERATION, GOOD SYSTEM
PERFORMANCE AND COOLING PURPOSES.
NOTICE
User only power supply cords, cables and plugs provided by or designated by GE Healthcare.
NOTICE
After turning off the system, wait at least 10 seconds before turning it on again. The system may not
be able to boot if power is recycled too quickly.

Table of Contents

Other manuals for GE LOGIQ E8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E8 and is the answer not in the manual?

GE LOGIQ E8 Specifications

General IconGeneral
BrandGE
ModelLOGIQ E8
CategoryMedical Equipment
LanguageEnglish

Related product manuals