GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
7 - 86 Section 7-7 - Troubleshooting Trees, Instructions and Tech Tips
7-7-23 Electromagnetic Interference (EMI) Troubleshooting
7-7-23-1 Prevention/abatement
For EMI rules and details, see "EMI Limitations" on page 2-4.
7-7-23-2 Different power outlet
Connect the LOGIQ™ S8 to another power outlet and check to see if the noise changes or disappears.
NOTE: Image artifacts can occur if, at any time within the facility, the ground from the main facility's incoming
power source to the LOGIQ™ S8 is only a conduit.
See "Minimal floor plan suggestions" on page 2-9.
7-7-23-3 Different system
Try another LOGIQ™ S8 at the same location and look for the same noise. If the noise is present on
the new system too, the noise is most likely from an external source/equipment.
7-7-23-4 Different location
Move the LOGIQ™ S8 to another location and check to see if the noise changes or disappears. This
may help to locate an external noise source.
Try to move the LOGIQ™ S8 to:
another location inside the room
another room
another floor
7-7-23-5 Disconnect external cables
Disconnect all external cables (network, all unused probes, ECG leads, etc.) and check to see if the
noise disappears.

Table of Contents

Other manuals for GE LOGIQ E8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E8 and is the answer not in the manual?

GE LOGIQ E8 Specifications

General IconGeneral
BrandGE
ModelLOGIQ E8
CategoryMedical Equipment
LanguageEnglish

Related product manuals