GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
2 - 4 Section 2-2 - General console requirements
2-2-3 EMI Limitations
Ultrasound systems are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transients in the air or wiring. Ultrasound systems also generate EMI. The LOGIQ
S8 complies with limits as stated on the EMC label. However, there is no guarantee that interference
will not occur in a particular installation.
Possible EMI sources should be identified before the unit is installed.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect. These
sources include:
Medical lasers
Scanners
Cauterizing guns
•Computers
•Monitors
•Fans
Gel warmers
Microwave oven
Light dimmers
Mobile phones
In-house wireless phones (DECT phones)
Wireless computer keyboard and mouse
Air conditioning system
High frequency (HF) surgery equipment
General AC/DC adapters
The presence of a broadcast station or broadcast van may also cause interference.

Table of Contents

Other manuals for GE LOGIQ E8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E8 and is the answer not in the manual?

GE LOGIQ E8 Specifications

General IconGeneral
BrandGE
ModelLOGIQ E8
CategoryMedical Equipment
LanguageEnglish

Related product manuals