GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
2 - 6 Section 2-2 - General console requirements
2-2-4 Probe Environmental Requirements
Probes should be operated, stored, or transported within the parameters outlined below.
NOTE: See attached TEE probe user manual for the environmental requirements of 6Tc-RS probe.
NOTE: See attached FibroScan user manual for the environmental requirements of FibroScan probe.
NOTE: Refer to Table 3-2 Acclimate Time on page 3-4 to determine the needed settlement time.
2-2-5 Time and Manpower Requirements
Site preparation takes time. Begin Pre-installation checks as soon as possible. If possible, allow six
weeks before delivery, for enough time to make necessary changes.
Table 2-4 Probe Environmental Requirements
Probe Operation Storage Transport
Temperature 2D probe
10° - 35° C/50° - 95° F
-40° - 60° C/-40° - 140° F
10C-D -10°- 55° C/14°-131° F-40° - 55° C/-40° - 131° F
4D probe 18° - 30° C/64.4° - 86° F-10° - 50° C/14° - 122° F
Humidity 2D probe 10 - 90% non-condensing
4D probe 30 - 75% non-condensing 10 - 85% non-condensing
Pressure 2D/4D probe 700 - 1060hPa
10C-D 680 - 1060hPa
NOTICE
!! NOTICE:
WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE PRODUCT SHOULD BE KEPT
IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
CAUTION
!! CAUTION:
Have two people available to deliver and unpack the LOGIQ S8.
Attempts to move the unit considerable distances (or on an incline) by one person alone, could
result in personal injury and/or damage to the system.