GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
1 - 16 Section 1-4 - Safety Considerations
WARNINGWARNING
DO NOT SUBSTITUTE PARTS OR MODIFY EQUIPMENT
BECAUSE OF THE DANGER OF INTRODUCING ADDITIONAL HAZARDS, DO NOT
INSTALL SUBSTITUTE PARTS OR PERFORM ANY UNAUTHORIZED MODIFICATION
OF THE EQUIPMENT.
WARNINGWARNING
ENSURE THAT THE SYSTEM IS TURNED OFF AND UNPLUGGED.
WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE AS THERE ARE NO
TEST POINTS TO VERIFY ISOLATION. THE BLUE LIGHT ON THE POWER BUTTON WILL TURN
OFF.
WARNINGWARNING
RISK OF ELECTRICAL SHOCK, SYSTEM MUST BE TURNED OFF AND
DISCONNECTED FROM POWER SOURCE. CORD MUST BE CONTROLLED AT ALL
TIMES.
WARNINGWARNING
USE EXTREME CAUTION AS LONG AS THE LOGIQ S8/LOGIQ E8 IS UN-STABLE,
NOT RESTING ON ALL FOUR CASTERS.
WARNINGWARNING
TILTING THE CONSOLE REQUIRES TWO PEOPLE IN ORDER TO AVOID INJURY TO
SERVICE PERSONNEL AND DAMAGE TO THE EQUIPMENT.
WARNINGWARNING
BEWARE OF POSSIBLE SHARP EDGES ON ALL MECHANICAL PARTS. IF SHARP
EDGES ARE ENCOUNTERED, THE APPROPRIATE PPE SHOULD BE USED TO
REDUCE THE RISK OF INJURY.