GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 7-4 - Service Desktop (CSD) - R1 to R4 7 - 21
7-4-3 Global Service User Interface (GSUI)
Internationalization
The user interface provided by the service platform is designed for GE personnel and is in English only.
There is no multi-lingual capability built into the Service Interface.
Service Login
1.) Local Access (R1 to R4) via Utility - Service.
2.) Service Login screen is displayed.
3.) As soon as the Common Service Desktop (CSD) is started, the Service [Home] Page appears
Access / Security
The service interface has different access and security user levels. Each user is only granted access to
the tools that are authorized for their use.
Every access request, whether successful or not, will be logged into a service access log that is
viewable to authorized users.
Restart LOGIQ S8 after diagnostics
Always shutdown the system and reboot after a diagnostics session. A red message on the task bar will
remind you that the system needs to be restarted after a diagnostic run or disruptive mode is enabled,
before it is returned for customer use.
Table 7-1 Access Authorization
USER LEVEL ACCESS AUTHORIZATION PASSWORD
Operator
Authorized access to specified diagnostics, error logs and utilities.
Same acquisition diagnostic tests as GE Service.
uls
Administrator uls
External Service gogems
GE Service Knowledge of the service level password.
rotating security
password