DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 7-7 - Troubleshooting Trees, Instructions and Tech Tips 7 - 85
7-7-22-2 Noise picked up from the air
7-7-22-3 Noise received through the external cables
Electromagnetic Interference (EMI) from radio frequencies, magnetic fields, and transients in the wiring.
The noise can enter the system through the mains power cable, probe cable(s) or any other external
connected cable(s).
Disconnect cables that are not needed for the basic use of the LOGIQ™ S8:
Network cable
Cables to any external peripherals
ECG cables and other cables connected to the Patient I/O
Verify whether the noise changed or disappeared when the cables were removed.
Often, this type of noise is due to grounding problems in the mains power system or when the LOGIQ™
S8 is sharing a power line with other equipment.
7-7-22-4 Intermittent noise
Intermittent noise can be caused by other equipment.
Table 7-17 Noise Conditions Picked Up from the Air
Symptom Cause Solution
Noise is coherent -“penlight noise”
pointing down in the picture due to the
fact that the noise is received on all
channels.
The electromagnetic
interference (EMI) from
radio frequencies,
magnetic fields, and
transients in the air.
Test whether the noise is picked up by a probe cable.
Noise is a problem on a single probe. Test with another probe.
Noise is a problem on a single probe
port.
Move the LOGIQ™ S8 to another location to see if the noise
persists.
Table 7-18 Noise Conditions for Intermittent
Symptom Solution
Intermittent noise Check equipment that is turned on and off near the scanner.
Intermittent noise present around the clock Check the clock.
Intermittent noise present at certain times
Check equipment that is turned on and off near the scanner at
certain times.

Table of Contents

Other manuals for GE LOGIQ E8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E8 and is the answer not in the manual?

GE LOGIQ E8 Specifications

General IconGeneral
BrandGE
ModelLOGIQ E8
CategoryMedical Equipment
LanguageEnglish

Related product manuals