GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 7-5 - Gathering Troubleshooting Data (R4.2.5x and later) 7 - 39
7-5-3 Shortcut keys
There are several shortcut keys on the LOGIQ™ S8 that can be useful during troubleshooting:
Table 7-3 Shortcut Keys
Key Description
Alt+1 or Alt+2 Marks events in the log file. Used for the flagging issues during operation.
Alt+F10
Displays SW Shut Down. If the LOGIQ™ S8 can be shut-down by using these keys, it
means that SW is working and the problem could be the ON/OFF switch or the wire that
carries the shut-down signal.
Alt+F5
For the echo simulator (keyboard simulation), brings up or removes the front panel
simulator. It may be used with VCO to press buttons on the front panel that are not standard
keys on your PC. .
F4 Displays the DICOM Job Spooler.
Alt+A Displays a cursor (Alt+A stands for arrow).
Alt+D Captures scanner logs.