GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 9-4 - Maintenance task schedule 9 - 5
Battery
Battery deterioration
When the system detects the battery deterioration, the
“Battery deterioration” dialog displays.
Replace the battery for proper function.
When “Battery
deterioration” message will
be displayed, call to GE
Service.
Battery Refresh
If the user doesn’t maintain the battery as indicated in
Basic User Manual Chapter 3 “Refreshing battery”,
refresh the battery every 6 months for the user.
Refresh procedure:
1. Turn on the system.
2. Wait until the battery is fully charged. it takes at least 1
hour to fully charge the battery.
3. Turn off the system.
4. Remove all probes.
5. Turn on the system.
6. Unplug the AC cable and wait until the system shuts
down. It may take 30 minutes or more to complete
shutdown.
7. Wait at least 5 hours.
8. Plug in the AC cable.
9. Turn on the system.
10. Wait until the battery is fully charged. It takes about 3
hours to fully charge the battery.
Every 6 months
Table 9-1 Customer Care Schedule
Item Service at Indicated Time D W M A Notes

Table of Contents

Other manuals for GE LOGIQ E8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E8 and is the answer not in the manual?

GE LOGIQ E8 Specifications

General IconGeneral
BrandGE
ModelLOGIQ E8
CategoryMedical Equipment
LanguageEnglish

Related product manuals