GE HealthCare CONFIDENTIAL
Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
4 - 8 Section 4-2 - General Procedures
4-2-3 Power ON/Boot Up (continued)
Connect AC (mains) power to the LOGIQ Totus
Connecting AC power to the LOGIQ Totus involves preliminary checks of the power cord, voltage level,
and compliance with electrical safety requirements.
NOTE: The LOGIQ Totus will function on voltages from 100-240 volts and 50 or 60 Hz. However, if
using 220 volt power, then a center tapped power source is required (North America Only).
1.) Ensure that the wall outlet is of appropriate type, and that the circuit breaker is turned off.
2.) Uncoil the power cable, allowing sufficient slack so that the LOGIQ Totus can be moved slightly.
3.) Verify that the:
- power cable is without any visible scratches, any sign of damage or tangled.
- on-site mains voltage is within the limits indicated on the rating label near the circuit breaker
on the rear of the LOGIQ Totus.
4.) Connect the female plug of the power cable to the power inlet at the rear of the LOGIQ Totus.
5.) Verify that the mains power circuit breaker is in OFF position, If not, switch it to OFF.
6.) Connect the other end (male plug) of the power cable to a hospital grade mains power outlet with
the proper rated voltage, and the LOGIQ Totus is ready for power ON/boot up.
Figure 4-2 Circuit Breaker