DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 124 Section 8-7 - Replacing OPIO
8-7-26 Replacement of the UHUB-T
This table includes information specific to these instructions. For more information 8-2-6 "Tools needed
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6.
8-7-26-1 Preparations
1.) Power off/Shutdown the system as described in 1-7-2 "Lockout/tagout procedure" on page 1-25
2.) Remove the OPIO Top Cover Set-T as described in 8-7-2 "Replacement of the OPIO Top Cover
Set-T" on page 8-89
3.) Remove the UOPIO-T Assy as described in 8-7-10 "Replacement of the UOPIO-T Assy" on page
8-98
8-7-26-2 Removal Procedure
8-7-26-3 Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8-
12 on page 8-223
NOTE: The UBE To OPIO HUB USB 3.0 Cable can be inserted to any of two USB ports (OPIO HUB).
Table 8-62 Manpower, Time, Tools, and PPE
Manpower Time Tools PPE
One person Approximately 15 minutes Phillips screwdriver
Cut resistant gloves
Knee Pad
Steps Corresponding Graphic
1. Unscrew 2 screws and unlock the Cable Clamp.
2. Disconnect cables which are connected to the
UHUB-T.
3. Unscrew the 12 screws to remove the UHUB-T from
the console.