GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 70 Section 8-5 - Replacing Cables
8-5-13 Replacement of the OPIO HUB To Monitor USB 2.0 Cable
This table includes information specific to these instructions. For more information 8-2-6 "Tools needed
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6.
8-5-13-1 Preparations
1.) Power off/Shutdown the system as described in 1-7-2 "Lockout/tagout procedure" on page 1-25
2.) Remove the T1 Monitor Arm Cover Set as described in 8-6-2 "Replacement of the T1 Monitor Arm
Cover Set" on page 8-74
3.) Remove the OPIO Top Cover Set-T as described in 8-7-2 "Replacement of the OPIO Top Cover
Set-T" on page 8-89
4.) Remove the UOPIO-T Assy as described in 8-7-10 "Replacement of the UOPIO-T Assy" on page
8-98
8-5-13-2 Removal Procedure
Table 8-31 Manpower, Time, Tools, and PPE
Manpower Time Tools PPE
One person Approximately 15 minutes
Phillips screwdriver
Nipper
Cut resistant gloves
Knee Pad
Steps Corresponding Graphic
1. Remove the Monitor Cable Cover.
2. Unscrew 2 screw and Remove the plastic clamp from
OPIO HUB To Monitor Power Cable and disconnect
the OPIO HUB To Monitor Power Cable from the
Monitor.
Note: If the HDU 23.8 inch Monitor LOGIQ Exx was
installed, remove the HDU 23.8 inch Monitor LOGIQ
Exx as described in
8-6-7 "Replacement of the
HDU 23.8 inch Monitor LOGIQ Exx" on page 8-
86
3. Cut 4 Cable Ties.