DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 84 Section 8-6 - Replacing Monitor and Monitor Arm
8-6-6 Replacement of the Main ULTRASOUND BEZELLESS 23.8INCH LCD T2
This table includes information specific to these instructions. For more information 8-2-6 "Tools needed
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6.
8-6-6-1 Preparations
1.) Power off/Shutdown the system as described in 1-7-2 "Lockout/tagout procedure" on page 1-25
8-6-6-2 Removal Procedure
Table 8-36 Manpower, Time, Tools, and PPE
Manpower Time Tools PPE
One person Approximately 15 minutes Phillips screwdriver
Cut resistant gloves
Knee Pad
Steps Corresponding Graphic
1. Remove the Monitor Cable Cover.
2. Unscrew 5 screws to remove the plastic clamp from
the console. Disconnect the HDMI Cable, Power
cable and USB Cable.
3. Unscrew 4 screws to remove the Main monitor from
the console.